Equipment Status
On this page, you can find the current status of equipment used at MCh, and its related issues as reported by responsible personnel.
Status Overview
Known Issues
Date | Equipment | Issues |
---|---|---|
09.09.2024 | V-UHV furnace | detected leak cause: being investigated status: not working, situation is being addressed |
06.09.2024 | DSC water deionizer | water supply turned off due to a water leak a bucket put beneath the deionizer to capture any dripping water status: water turned back on, no leak found |
03.09.2024 | GADDS | instrument is not operational cause: unknown status: not working, contacting the manufacturer |
15.08.2024 | GADDS | display of a newly installed second-hand manual controller not working properly status: working, but refer to the manual |
23.07.2024 | SEM JEOL | experiencing evacuation-related error, issue usually resolved upon restarting possible cause: improperly positioned door seal status: working, but need to pay attention to errors |
11.07.2024 | compressed air compressor | compressor not working, IOB working on its repair cause: broken air condition, warm weather, extensive amount used by IOB for experiments status: working, but issue could repeat |